Recall Z-0021-2014 ยท OWB Interventional Fluoroscopic X-Ray System

GE Healthcare, LLC

GE Healthcare, Advantx, Innova 2000, Innova 2100IQ, Innova 3100, Innova 3100IQ, Innova 4100, Innova 4100IQ. Intended for general purpose diagnostic fluoroscopy and radiography studies.

DateOct 17, 2013 (initiated Aug 16, 2013)
StatusTerminated
ReasonGE Healthcare has become aware of a potential safety issue involving missing screws in the overhead Video Monitor Suspension of the Advantx, Innova 2000, Innova 2100IQ, Innova 3100, Innova 3100IQ, Innova 4100, Innova 4100IQ Cardiovascular X-ray imaging systems.
Root causeProcess control
ActionConsignees were sent on 8/12/2013 a GE Healthcare "Urgent Medical device Correction" letter dated August 12, 2013. The letter was addressed to Hospital Administrators / Risk Mangers, Mangers of Radiology /Cardiology /Surgery Department and Radiologists /Cardiologists /Surgeons. The letter described the Safety Issue, safety Instructions, Affected Product details, Product Correction and Contact Information (US 800-437-1171, Japan 0120 055 919).
Quantity2184 (1063 US. 1121 OUS)
DistributionWorldwide Distribution-USA (nationwide) including DC and Puerto Rico, and except ND, SD and VT, and the countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELARUS, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, COSTA ARICA, CROATIA, CZECH REPUBLIC, DOMINICAN REPUBLIC, ECUADOR, EGYPT, FI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022322INNOVA AND INNOVA 2000 S
Ge Medical Systems, Inc., Oct 8, 2002
K023178INNOVA 4100
Ge Medical Systems, Inc., Nov 26, 2002
K945375ADVANTX LCA
GE Medical Systems, Jan 30, 1995
K960575ADVANTX LCV+
GE Medical Systems, May 7, 1996
K974367ADVANTX LCN, ADVANTX LCLP
GE Medical Systems, Feb 13, 1998
K993037LCV+ VERSION 2
Ge Medical Systems, Inc., Feb 14, 2000