Recall Z-0152-2014 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Beckman Coulter Inc.

ISE REF Valve, Part No. MU3226 for the AU400, AU640, AU2700, and AU5400 Chemistry Analyzer. Automated chemistry analyzers that measures analytes in samples.

DateNov 5, 2013 (initiated Sep 9, 2013)
StatusTerminated
ReasonThe recall was initiated because Beckman Coulter has confirmed that some AU Clinical Chemistry Analyzer ISE units can experience premature ISE REF (Reference) valve failure. Premature valve failure can be detected by the presence of air bubbles in the ISE REF solution tubing on the ISE unit. If air bubbles are present, accuracy of the ISE test results may be affected.
Root causeNonconforming Material/Component
ActionA customer notification letter dated 9/9/13 was sent to customers who received the ISE REF Valve, Part Number MU3226 for the AU400, AU640, AU2700, AU5400 and the ISE REF Valve, Part Number MU7638 for the AU480, AU680, AU5800 Chemistry Analyzers. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with questions are instructed to contact Beckman Coulter Customer Support Center via http://www.beckmancoulter.com/customersupport/support, via phone at (800) 85
Quantity4584 units total (1173 units in US)
DistributionWorldwide Distribution-USA (nationwide) and the countries of Canada, Vietnam, Zambia, and Zimbabwe.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K003721MODIFICATION TO OLYMPUS AU2700 CLINICAL CHEMISTRY ANALYZER
Olympus America, Inc., Dec 18, 2000
K011720OLYMPUS AU5400 CLINICAL CHEMISTRY ANALYZER
Olympus America, Inc., Jun 25, 2001
K961274OLYMPUS AU600 CLINICAL CHEMISTRY ANALYZER
Olympus America, Inc., Jul 5, 1996
K981743OLYMPUS AU400 CLINICAL CHEMISTRY ANALYZER
Olympus America, Inc., Jun 9, 1998