Recall Z-0010-2014 ยท BSZ Gas-Machine, Anesthesia

GE Healthcare, LLC

GE Healthcare, Avance, Avance CS2, Aisys, Aespire View, and Engstrom. Intended for volume or pressure control ventilation.

DateOct 15, 2013 (initiated Aug 28, 2013)
StatusTerminated
ReasonGE Healthcare has recently become aware of a potential safety issue involving unresponsive buttons on the display of the Engstrm ventilator and the Aespire View, Aisys, and Avance anesthesia machines. As a result of a manufacturing issue, the buttons on the left, right, and bottom keypads may not always detect button presses. This may result in the inability to access certain menu functions wh
Root causeNonconforming Material/Component
ActionConsignees were sent on 8/28/2013 a GE Healthcare "Urgent Medical device Correction" letter dated August 28, 2013. The letter was addressed to Chief of Anesthesia, Health Care Administrator / Risk Manager and director of Biomedical / clinical engineering. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information (1-800-345-2700 US).
Quantity911 (252 US, 659 OUS)
DistributionWorldwide Distribution-USA (nationwide) including the states of US: AK, AZ, CA, CO, DE, FL, GA, IL, KY, LA, MD, MI, MN, MS, MT, NV, NY, NC, OH, OR, PA, SC, TX, WY, and WI and the countries of AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHILE, CHINA, COST ARICA, CROATIA, CZECH R
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110213GE DATEX-OHMEDA AISYS
Datex-Ohmeda, Inc., Jul 20, 2011
K111116ENGSTROM CARESTATION, ENGSTROM PRO, MODELS 1505-9000-000, 1505-9003-000
Datex-Ohmeda, Sep 9, 2011
K112722GE DATEX-OHMEDA AVANCE
Datex-Ohmeda, Inc., Dec 16, 2011
K122445AESPIRE 7900, AESPIRE VIEW
Datex-Ohmeda, Oct 12, 2012
K123125GE DATEX-OHMEDA AVANCE CS2
Datex-Ohmeda, Inc., Feb 6, 2013