Recall Z-0168-2014 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Roche Diagnostics Operations, Inc.
COBAS INTEGRA 800 Analyzer with software 9864.C2 Cobas. COBAS Part Numbers 28122474001, 28122474692, 04559126001, 04589459970 corresponding to the Analyzer with or without closed tube sampling accessory A fully automated system for clinical chemistry analysis intended for the in vitro quantitative/qualitative determination of analytes in body fluids.
| Date | Nov 5, 2013 (initiated Sep 16, 2013) |
| Status | Terminated |
| Reason | Roche COBAS INTEGRA 800 system, Software Version 9864.C2 does not perform necessary ISE Service Actions. Whenever the ISE rack is removed, the ISE counters in the software are reset to zero and the corresponding Service Actions are not conducted automatically by the system. Additionally, certain maintenance activities that have to be performed manually by the operator are not requested on the sof |
| Root cause | Software Design Change |
| Action | Roche sent an Urgent Medical Device Correction letter dated September 16, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Roche Diagnostics recommends tracking all ISE counters and manually requesting the corresponding Service Actions listed in the notification be monitored until updated software becomes available during Q4 2013. If the affected product was moved to another site, the notification is to be forwarded. consignees were asked to complete the attached faxback form and fax it to 1-877-266-0997 and fil |
| Quantity | 527 units |
| Distribution | Nationwide Distribution and Puerto Rico |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K951595 | ROCHE COBAS INTEGRA ANALYZER, REAGENT CASSETTES Roche Diagnostic Systems, Inc., Sep 8, 1995 |