Recall Z-2273-2013 ยท CBK Ventilator, Continuous, Facility Use

Carefusion 211 Inc dba Carefusion

Product Name: AVEA Ventilator. All Models. The affected devices are cleared under K073069 pm 1/25/2008 and K103211 on 05/06/2011. This device is a Class II medical Device. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators

DateSep 27, 2013 (initiated Sep 5, 2013)
StatusTerminated
ReasonCareFusion has identified a potential risk associated with AVEA ventilators when used at higher altitudes. A proactive complaint review identified an error in the barometric pressure compensation. CareFusion is voluntarily initiating a field correction of the affected devices to correct the error.
Root causeComponent design/selection
ActionCareFusion sent customer notification letters on September 05, 2013 via USPS mail return receipt carrier. The customer notification letter dated September 3, 2013 was titled "URGENT PRODUCT RECALL", and was addressed to Biomedical Engineering, Respiratory Care, and Risk Management. The customer notification letter informed customers of the product under recall with product name, description, and serial numbers enclosed as list unique to each customer. The customer notification letter also states the Problem and Affected Devices: Issue, Potential Risk, Actions to be Taken by CareFusion, Action
Quantity19,546 units
DistributionWorldwide distribution: US (nationwide) including Puerto Rico and countries of: Albania, American Samoa, Argentina, Australia, Austria, Bahrain, Barbados, Bangladesh, Belgium, Bermuda, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Canada, Chile,China, Colombia, Costa Rica, Croatia, Cyprus, Czech