ClearanceBrief
Recall Z-0277-2014 ยท JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented

Zimmer, Inc.

VerSys Cemented Revision Calcar (CRC) Hip System Femoral Stems VerSys Heritage Hip System Revision Femoral Stems. Indicated for total hip arthroplasty.

DateNov 14, 2013 (initiated Aug 28, 2013)
StatusTerminated
ReasonBased on a review of historic packaging validations, it was determined that a subset of product packaged in one of the clean rooms at a single Zimmer facility required testing to ensure that previously packaged product from this facility met the established requirements for sterile barrier integrity. The number of samples required for packaging testing (seal integrity and seal strength) to provid
Root causePackaging process control
ActionRemoval of the subject devices from the field including distributor and hospital inventories. Distributors and Risk Managers and Surgeons at hospitals that received a direct shipment from Zimmer or its Distributors received a notification on 09/19/2013.
Quantity12866 units
DistributionWorldwide distribution: US (Nationwide) including Puerto Rico and Virgin Islands; and countries of: Australia, Algeria, Belgium, Bolivia, Canada, Canary Islands, Chile, Croatia, Finland, France, Germany, Greece, Guatemala, Honduras, India, Israel, Italy, Japan, Jordan, Lima-Peru, Malaysia, Morocco,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K913649MODULAR CALCAR REPLACEMENT
Zimmer, Inc., Nov 13, 1991
K963109HERITAGE HIP SYSTEM
Zimmer, Inc., Jan 9, 1997