Recall Z-0026-2014 ยท LWP Implantable Pulse Generator, Pacemaker (Non-Crt)

Boston Scientific CRM Corp

Boston Scientific TELIGEN, Implantable Cardiac Defibrillator models E102, E110, the following models are not available in the US: models F102, F110, F111. Sterilized using ethylene oxide. Boston Scientific, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. Indicated for patients who are at risk for sudden cardiac death due to ventricular arrhythmias and may require pacing support.

DateOct 17, 2013 (initiated Aug 29, 2013)
StatusTerminated
ReasonThe performance of a low voltage capacitor in a subset of COGNIS CRT-Ds and TELIGEN ICDs manufactured prior to December 2009, may be compromised over time, causing increased current drain that can lead to premature battery depletion.
Root causeComponent design/selection
ActionAn Important Medical Device Information letter dated August 2013 was sent to implanting and following physicians. The letter provided a description of the issue, describes clinical implications, provides the rate of occurrence, and gives recommendations for monitoring patients.
Quantity38,500 total worldwide
DistributionWorldwide Distribution-USA (nationwide) including Guam and Puerto Ricoand the countries of Argentina, Aruba, Australia, Austria, Belgium, Bermuda, Brazil, Canada, Chile, Hong Kong, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Finland, France, Gabon, Germany, Gr
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.