Recall Z-2253-2013 ยท OWB Interventional Fluoroscopic X-Ray System

GE Healthcare, LLC

GE Healthcare Discovery IGS 730 Cardiovascular X-Ray Imaging System. The angiography X-ray systems are indicated for use in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures. Additionally, with the OR table, the angiographic X-ray systems are indicated for use in generating fluoroscopic and rotati

DateSep 18, 2013 (initiated Jun 19, 2013)
StatusTerminated
ReasonGE Healthcare has recently become aware of a potential safety issue with respect to Discovery IGS 730 Imaging Systems. A patient data management error may prevent the Discovery IGS 730 Imaging System from booting on start-up or re-booting during procedure. The boot issue may occur when a large amount of patient data remains in the browser and is not deleted. The inability to boot the system ma
Root causeSoftware design
ActionThe firm, GE Healthcare, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated June 19, 2013 to its Consignees/customers. The letter described the Safety Issue, Safely Instructions, Affected Product Detail, Product Correction and Contact Information. The consignees were instructed to please ensure that all potential users in your facility are made aware of this safety notification and the recommended actions. GE Healthcare will correct all affected systems at no cost to you. A GE Healthcare service representative will contact you to arrange for this correction. If you have any questions
Quantity4 (2 US, 2 OUS)
DistributionWorldwide distribution: US (nationwide) including states of : PA and TN; countries of: FRANCE and SINGAPORE.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113403GE DISCOVERY IGS ANGIOGRAPHIC,FLUOROSCOPIC X-RAY SYSTEM
GE Medical Systems SCS, Feb 14, 2012