Recall Z-2191-2013 ยท GEX Powered Laser Surgical Instrument

Lumenis, Inc.

Vision One Laser System Model GA-0025020, Serial No: 10050, 10051, 10053. Intended for use in the treatment of ocular pathology.

DateSep 11, 2013 (initiated Aug 19, 2013)
StatusTerminated
ReasonLumenis has initiated a recall on certain models of Vision One System due to a potential for unintended laser exposure to the user.
Root causeSoftware design
ActionCustomer's were notified via letter on 8/19/13. Service visits to consignees were scheduled with anticipated completion on 8/26/13 to replace the control board. Consignee monitoring was performed by use of return reply verification tracking cards returned by the Lumenis service engineer after completion of CPU board replacement.
Quantity3 units
DistributionDistributed in the states of NJ, PA, and IL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111213VISION ONE LASER SYSTEM
Lumenis, Inc., Jul 25, 2011