Recall Z-0065-2014 ยท JAQ System, Applicator, Radionuclide, Remote-Controlled

Varian Medical Systems, Inc.

VARISOURCE Breast Template System for high-dose rate Brachytherapy. ***INSTRUCTIONS FOR USE*** Breast Bridge & Template System AL13011000 ***Manufacturer: Varian Medical Systems, Inc., 3100 Hansen Way, Palo Alto, CA 94304 USA

DateOct 24, 2013 (initiated Aug 9, 2013)
StatusTerminated
ReasonThe recall was initiated after Varian Medical became aware the Lexan templates of the Breast Bridge Template System could become bent and turn opaque after steam sterilization.
Root causeNonconforming Material/Component
ActionUrgent Medical Device Correction, Urgent Field Safety Notice were mailed by FedEx tracked delivery to all consignees on August 12, 2013.
Quantity3 sets
DistributionWorldwide distribution: US (nationwide) distribution in CA and in the country of India.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023220VARISOURCE BREAST TEMPLATE SYSTEM
Varian Medical Systems, Oct 25, 2002