Recall Z-0179-2014 ยท IZL System, X-Ray, Mobile
Carestream Health Inc.
DRX- Revolution, Catalog number: 101937, 1023415 & 1023423. Made in U.S.A. by Carestream Health, Inc. 150 Verona Street Rochester, NY 14608 to perform radiographic x-ray examinations on all pediatric and adult patients
| Date | Nov 7, 2013 (initiated Aug 29, 2013) |
| Status | Terminated |
| Reason | Carestream Health Inc. has conducted a voluntary field action which includes the installation of a software patch on the following: DRX-Revolution Mobile X-Ray, DRX-Mobile and DRX Transportable Systems with DirectView Version 5.5 and 5.6 console software when used with DRX-1 Detector version 69 firmware. Carestream recently received reports of a non-diagnostic image on exams when a Normal Exposure |
| Root cause | Software design |
| Action | The firm, Carestream Health Inc., sent an "Urgent: Medical Device Recall" letter dated August 29, 2013 to its customers. The letter described the product, problem and actions taken. Carestream has developed a software patch that will allow Normal Exposure Mode and Extended Exposure Mode images to be taken in any order and will install the patch in all DRX-Revolution, DRC-Mobile and DRX-Transportable Systems with software version 5.5 or higher and Detector firmware version 69 or higher. No DRX Transportables were affected. Your Carestream service representative will install the patch at your |
| Quantity | 190 units |
| Distribution | Nationwide distribution:AL, AZ, AR, CA, CO, CT, DC, FL, GA, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MI, MO, NE, NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, UT, VT, VA, WA, WI, and WV. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K120062 | DRX-REVOLUTION MOBILE X-RAY SYSTEM Carestream Health, Inc., Apr 13, 2012 |