ClearanceBrief
Recall Z-2233-2013 ยท MAI Fastener, Fixation, Biodegradable, Soft Tissue

Smith & Nephew, Inc. Endoscopy Division

TWINFIX Ultra PLLA/HA 5.5 mm Suture Anchor with 2 ULTRABRAID Suture, White / Black: Part Number: 72202602 Biodegradable suture anchor

DateSep 17, 2013 (initiated Aug 6, 2013)
StatusTerminated
ReasonPinholes detected in a small number of primary pouches that may constitute a breach of the sterile barrier.
Root causePackaging
ActionSmith&Nephew sent an Urgent Product Recall Notification letter dated August 6, 2013, via Federal Express to all affected customers. Addendum customer letters for 4 additional part numbers were sent via Federal Express on August 19, 2013, The letters identified the product, the problem, and the action to be taken by the customer. Customers were asked to inspect their inventory, locate unused devices and quarantine them. Customers were also asked to complete the last three columns in the inventory Return Certification form indicating the quantities that need to be returned, and include the
Quantity7148 US
DistributionWorldwide Distribution - USA (nationwide) Canada, Europe: Austria, Belgium, Cyprus, Czech Republic, Denmark, Germany, Greece, Finland, France, Italy, Ireland, Luxembourg, Netherlands, Norway, Poland, Portugal, Russia, Slovenia, Spain, Sweden, Switzerland, Turkey, Ukraine and United Kingdom, Asia-pa
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K093844SMITH & NEPHEW TWINFIX ULTRA HA SUTURE ANCHOR
Smith & Nephew, Inc., Apr 1, 2010