Recall Z-2173-2013 ยท MAQ Kit, Dna Detection, Human Papillomavirus

Roche Molecular Systems, Inc.

KIT cobas 4800 HPV AMP/DET 240T / 960T US-IVD, CE-IVD Roche Molecular System, Inc. 1080 US highway 202 South Branchburg, NJ 08876 USA Roche Diagnostics 9115 Hague Road Indianapolis, IN 46250-0457 USA The cobas HPV Test is a qualitative in vitro test for the detection of Human Papillomavirus in patient specimens. The test utilizes amplification of target DNA by the Polymerase Chain Reaction (PCR)

DateSep 10, 2013 (initiated Jul 25, 2013)
StatusTerminated
ReasonDuring an internal investigation, a formulation error that occurred was identified during the manufacture of the cobas 4800 HPV Master Mix reagent that was packaged in four (4) cobas 4800 AMP/DET kits, US- and CE-IVD (240 and 960 test kit configurations). Examination of the batch production record of the material revealed there was an error in the tare weight used for the carboy during the fina
Root causeDevice Design
ActionRoche Molecular Diagnostics sent an Urgent Medical Device Correction letter dated July 25, 2013, to all affected customers via UPS return receipt. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to: 1. Immediately discontinue use of the product lots listed above. 2. Discard affected product according to local guidelines. 3. To receive replacement inventory, complete the entire attached faxback form and fax it to 1-866-918-8741. 4. The faxback form must be completed and returned even if you do not have affected invent
QuantityUS-IVD: 350 Kits; CE-IVD: 441 KITS
DistributionWorldwide Distribution - USA (nationwide) including the countries of Austria, Australia, Belgium, Brazil, Czech Rep, Canada, Croatia, Columbia, Denmark, Dominican Republic, Germany, Ecuador, France, Japan, Greenland, Mexico, Italy, New Zealand, Netherlands, Peru, Norway, Singapore, Romania, South Af
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.