ClearanceBrief
Recall Z-2192-2013 ยท FMK Single Use Only Blood Lancet With An Integral Sharps Injury Prevention Feature

Becton Dickinson & Company

BD Microtainer Quikheel Lancet 1.0 mm incision Depth 2.5 mm Incision Length REF 368101 Rx Only Becton, Dickinson and Company, Franklin Lakes, NJ 07417 USA Made in USA. Used to perform heel sticks for the purpose of collecting skin puncture blood.

DateSep 11, 2013 (initiated Jul 10, 2013)
StatusTerminated
ReasonThe Lancet will not retract after activation thus creating a potential for exposure to a used sharp.
Root causeComponent change control
ActionBD notified their customers and distributors by phone, e-mail, and/or letter/notice of return/ customer packing list on 7/10/2013.
QuantityUS - 39,800 units ; Canada - 18,400 units
DistributionDistributed in the states of CT, KY, IL, NY, PA, TX, and VA and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall