Recall Z-2115-2013 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

GE Healthcare, LLC

GE Healthcare, Dash 3000, Dash 4000 and Dash 5000. Product Usage: The intended use of the system is to monitor physiologic parameter data on adult, pediatric and neonatal patients.

DateAug 28, 2013 (initiated Apr 3, 2013)
StatusTerminated
ReasonGE Healthcare has recently become aware of a potential safety issue due to intermittent noise on ECG and/or Respiration waveforms associated with movement of the ECG connector on your DASH, PDM, and/or Tram Module. ECG and Respiration signal noise caused by DASH, PDM, and/or Tram ECG cable connector movement may result in reduced ECG, Respiration and arrhythmia detection performance.
Root causeNonconforming Material/Component
ActionGE Healthcare "Urgent Medical Device Correction" letter dated April 3, 2013 to all affected customers. The letter was addressed to Healthcare Administrator / Risk manger, Chief of Nursing and Director of Biomedical Engineering. The letter described the Safety Issue, Product, Safety Instructions, Affected Product Details, and Product Correction. For questions contact Technical Support at 1-800-558-7044.
Quantity28,120 *14,709 US & 13,411 OUS)
DistributionWorldwide Distribution - USA Nationwide including DC and PR except ND and the countries of: ALGERIA, ANGOLA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHAMAS, BAHRAIN, BANGLADESH, BELARUS, BELGIUM, BOSNIA-HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CZECH REPUBLIC, DENMARK, ECUAD
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K073462MODIFICATION TO DASH 3000, 4000, 5000 MONITOR AND ACCESSORIES
Ge Medical Systems Information Technologies, Jan 11, 2008