Recall Z-0114-2014 ยท JDQ Cerclage, Fixation

Synthes USA HQ, Inc.

Synthes Application Instrument Sternal Zip Flex System intended for use in primary or secondary closure/repair of the sternum following sternotomy or fracture of the sternum to stabilize the sternum and promote fusion

DateOct 30, 2013 (initiated Aug 13, 2013)
StatusTerminated
ReasonWhen using the 1st generation Zip Fix Application Instrument it is possible to cut the implant while the tensioning trigger is being squeezed which could lead to loosening of the implant.
Root causeLabeling False and Misleading
ActionSynthes sent a Urgent Medical Device Notification dated August 13, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Action to be taken: This notification is to make you aware of the issue and the importance of completely releasing the tensioning trigger before cutting the implant. Please take the following action: "To ensure your receipt and understanding of this information, Please complete the verification section of this letter and return to Synthes by either of the following methods: " Fax: 1-888-352-9616
Quantity319
DistributionWorldwide Distribution - US (nationwide) and country of Mexico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110789SYNTHES STERNAL ZIPFIX SYSTEM
Synthes (Usa), Jul 28, 2011