Recall Z-2276-2013 ยท MUJ System, Planning, Radiation Therapy Treatment

Elekta, Inc.

Monaco RTP System Radiation treatment planning

DateSep 25, 2013 (initiated Jul 2, 2013)
StatusTerminated
ReasonUpgrade installation script is designed to overwrite the same pouch files in the Treatment Couch Library with the original default values.
Root causeDevice Design
ActionMonaco 3.3 upgrades have been temporarily halted and Important Safety Notice (LINMON0007) has been issued to effected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to ensure that couch files are checked for releases 3.20.00 and 3.30.00 to verify that the correct relative electron density values are being used in the dose calculation. If customers have edited the sample couch files, and previously upgraded to Monaco 3.20.00 or 3.30.00, their couch files have already been overwritten. Open the couch files, mak
Quantity45
DistributionNationwide Distribution including AK, CA, CO, CT, FL, IL, IN, MI, MN, MO, NC, NJ, NY, OH, OK, OR, PA, SC, TN, TX, WA and Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110730MONACO RTP SYSTEM
Computerized Medical Systems, Inc., Jun 24, 2011