Recall Z-2266-2013 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Healthcare, LLC

GE 1.5T Signa HDxt, 3.0T Signa HDxt, Signa Vibrant, HD23 Software and the following model numbers: 2395001, 2395001-2, 2395001-3, 2395001-4, 5269997-2, 5269997-4, 2377062-61, 5127452-3, 5127452, 5337894-3, 5337894-4, 5911000-4, 5911000-5, 53085735308573-2. GE Discovery MR750w 3.0T, Discovery MR750 3.0T, Optima MR450w 1.5T, Discovery MR450 1.5T. The 1.5T Signa HDx family and 3.0T Signa HDx famil

DateSep 23, 2013 (initiated Apr 9, 2013)
StatusTerminated
ReasonGE Healthcare has recently become aware of a potential safety issue due to non-unique series identifiers embedded in image headers of certain GE MR Products. For ClariView and Screen Save from GVTK (Gems Visualization ToolKit) Viewer, the MR Product can generate image headers with non-unique series identifiers, which when exported to some PACS systems, results in the series being associated wit
Root causeSoftware design
ActionCustomers were sent two GE Healthcare "Urgent Medical Device Correction" letters one dated April 9, 2013, and the other May 22, 2013, each addressing one of the 2 groups of MR Products involved in the recall. The letters were addressed to Hospital Administrators / Risk managers, Radiology Department Mangers and Radiologists. The letters described the Safety Issue and provided Safety Instructions, Affected Product Details, Product Correction and Contact Information. GE Healthcare will correct all affected systems by providing a software update at no cost to you. A GE Healthcare service r
Quantity1,246 (521 US and 725 OUS)
DistributionWorldwide Distribution - All states in continental USA including DC except AK, AZ, RI and WY. and the countries of : AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, CZECH REPUBLIC, DENMARK, EGYPT, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, INDIA,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081028SIGNA 3.0T MR750 SYSTEM
Ge Medical Systems, LLC, Apr 25, 2008
K083147DISCOVERY MR450
Ge Medical Systems, LLC, Nov 4, 2008
K103327DISCOVERY MR750W 3.0T
Ge Healthcare Japan Corporation, Sep 30, 2011
K113490OPTIMA MR450W
Ge Medical Systems, LLC, Jan 13, 2012
K1216761.5T SIGNA HDXT 3.0T SIGNA HDXT SIGNA VIBRANT
Ge Medical Systems, LLC, Sep 20, 2012