Recall Z-0426-2014 ยท KOG

Custom Ultrasonics, Inc.

Custom Ultrasonics System 83 Plus 2 and Plus 9 Endoscope Washer/Disinfector.

DateNov 27, 2013 (initiated Mar 12, 2013)
StatusTerminated
ReasonCertain System 83 Plus 2 and System 83 Plus 9 units running the Windows 7 operating system were not fully validated prior to distribution.
Root causeSoftware design
ActionA notification letter, dated February 26, 2013, was sent to direct accounts to inform them of the recall and that technicians would visit facilities to replace the Windows 7 operating system with an appropriately validated Windows XP operating system.
Quantity118
DistributionWorldwide Distribution: USA (nationwide) and Internationally to: Canada, Slovakia Republic, Sydney and Peru.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122172SYSTEM 83 PLUS
Custom Ultrasonics, Jan 25, 2013
K983017MODIFICATION TO SYSTEM 83 PLUS
Custom Ultrasonics, Sep 22, 1998