Recall Z-2161-2013 ยท BSZ Gas-Machine, Anesthesia

GE Healthcare, LLC

Tec 6 continuous Flow Vaporizer and Tec 6 Plus Anesthesia Vaporizer. The Datex-Ohmeda Tec 6 Plus and Tec 6 Plus NAD Variant is an electronic vaporizer which delivers the anesthetic agent desflurane. The Tec 6 Plus attaches only to Datex-Ohmeda anesthesia machines equipped with the proprietary Selectatec manifold. The Tec 6 Plus NAD Variant attaches only to NAD anesthesia machines equipped wit

DateSep 6, 2013 (initiated Jul 29, 2013)
StatusTerminated
ReasonGE Healthcare has recently become aware of a potential safety issue involving the pre-operative check, low pressure leak test at 1% dial setting of the Tec 6 and Tec 6 Plus Vaporizers. The user manual pre-operative check, low pressure leak test may not detect the full range of leaks from seal wear degradation in the vaporizers. If the leak is not detected, it will result in fresh gas leaking to
Root causeComponent change control
ActionGE Healthcare sent an Urgent Medical Device Correction letter dated August 1, 2013, to all affected customers. The letter described the issue, asked that the Tec 6 and Tec 6 Plus Vaporizers user manual be updated with the attached addendum. Also Users are to perform the user manual recommended pre-operative check which includes the low pressure leak test on every vaporizer with the dial turned to the 12% setting instead of the specified 1% setting as per the attached addendum. If the device does not pass the new low pressure leak at 12%, it should be removed from service and consignee shoul
Quantity31,443 devices (23,947 US, 7496 OUS)
DistributionWorldwide Distribution - USA (nationwide) Puerto Rico and internationally to AUSTRALIA, AUSTRIA, BAHAMAS, BAHRAIN, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHILE, CHINA, COSTA RICA, CROATIA local name-Hrvatska, CYPRUS, DENMARK, ECUADOR, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, JOR
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K000275DATEX-OHMEDA TEC 6 PLUS ANESTHESIA VAPORIZER, DATEX-OHMEDA TEC 6 PLUS NAD VARIANT ANESTHESIA
Datex-Ohmeda, Inc., Apr 12, 2000
K925580OHMEDA TEC 6 (CFV) CONTINUOUS FLOW VAPORIZER, NAD
Ohmeda Anesthesia Systems Therapy Group, Jun 2, 1993