Recall Z-0023-2014 ยท JAK System, X-Ray, Tomography, Computed

GE Healthcare, LLC

GE, Optima CT660. Intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data.

DateOct 17, 2013 (initiated Mar 26, 2013)
StatusTerminated
ReasonGE Healthcare has recently become aware of a potential safety issue related with the DMPR function and Screen Saved Images on your scanner which involves Optima CT660, Brivo CT385, Discovery PET/CT 610, Discovery PET/CT 710. The DMPR and Screen Save functionality may possibly generate the same Series UID. It has been seen that some PACS may combine series under a patient using only the Series UID
Root causeSoftware design (manufacturing process)
ActionConsignees were sent a GE Healthcare "Urgent Medical Device Correction" letter dated March 26, 2013. The letter was addressed to Director of Radiology, Emergency Medicine Physicians and Radiologic Technologists. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information.
Quantity380 (52 US & 328 OUS)
DistributionWorldwide Distribution-USA (nationwide) including the states of AZ, CA, CO, FL, GA, IL, IA, IS, LA, MA, MO, NH, NM, NY, OH, OK PA, SC, TX, UT, VA, WA, and WI, and the countries of UNITED KINGDOM, DENMARK, CHILE, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELGIUM, CANADA, CZECH REPU
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110227OPTIMA CT660
Ge Medical Systems, LLC, Jul 20, 2011