Recall Z-2131-2013 ยท JTC Device, Microtiter Diluting/Dispensing

Copan Italia

WASP (Walk Away Specimen Process). Automatic Planting and Streaking Instrument Catalog number: W086 WASP is a fully automated system used to process microbiological specimens in liquid phase for the investigation of infectious microbes.

DateAug 30, 2013 (initiated Aug 9, 2013)
StatusTerminated
ReasonThe recall of the WASP (Walk Away Specimen Process) instrument was initiated because Copan Italia SpA has upgraded the software concerning the broth inoculation table feature. The upgrade has been considered necessary from the manufacturer in order to eliminate the very low risk of rotation of the broth table when the door is open.
Root causeDevice Design
ActionCopan sent an Urgent WASP recall letter dated July 31, 2013, to all affected customers that received the WASP (Walk Away Specimen Process) instrument to notify them of the software upgrade concerning the broth inoculation table feature. The upgrade has been considered necessary from the manufacturer in order to eliminate the very low risk of rotation of the broth table when the door is open. This event may happen only if damage in the door locking system happens. This could result in a minor, transient injury to the user due to the impact with the rotating table. The WASP instruments involv
Quantity63 units
DistributionWorldwide Distribution - USA including Oregon and California., and the countries of France, Belgium, Italy, Denmark, Norway, Switzerland, Romania, Finland, Germany, South Africa, Canada, Ireland, UK, Sweden and Japan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.