Recall Z-2137-2013 ยท LIT Catheter, Angioplasty, Peripheral, Transluminal
Boston Scientific Corporation
Boston Scientific, Small Peripheral Cutting BalloonTM, Sterilized using irradiation: The Peripheral Cutting Balloon Catheters are indicated for Percutaneous Transluminal Angioplasty (PTA) of obstructive lesions of synthetic or native arteriovenous dialysis fistulae. 1) 4.00mm x 15mm Small Peripheral Cutting BalloonTM MonorailTM Microsurgical Dilatation Device, UPN #001BPM4015140F0, Catalogue #
| Date | Sep 3, 2013 (initiated Jul 31, 2013) |
| Status | Terminated |
| Reason | Boston Scientific is initiating a recall removal of the 4.0mm x 15mm size of Small Peripheral Cutting Balloon (Monorail and Over-the-Wire) Microsurgical Dilatation Devices and Flextome Cutting Balloon (Monorail and Over-the-Wire) Microsurgical Dilation Devices. Boston Scientific has determined that users may experience significant difficulty or inability to remove the protector cap from the devic |
| Root cause | Process change control |
| Action | Boston Scientific sent an Urgent Medical Device Recall Removal letter dated July 30, 2013, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were advised that further distribution or use of the product should cease immediately, and to segregate the product and return it to Boston Scientific. Consignees were asked to complete and return the "Recall Removal Reply Verification Tracking Form". For questions consignees were instructed to contact their local Sales Representative. For questions regarding this recall |
| Quantity | 1,331 |
| Distribution | Worldwide Distribution - USA including AL, AZ, AR, CA, CO, CT, DE, DC, FL, GA, ID, IL, IN, IA, KY, LA, MD, MA, MI, MN, MS, MO, MT, NE, NV,NH,NJ, NY, NC, ND, OH, OK, OR, PA, RI, SC,SC, TN, TX, UT, VA, WA, WV, and WI. Internationally to Algeria, Finland, Japan, Spain, Australia, France, Latvia, Swed |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.