Recall Z-2170-2013 ยท CGN Acid, Folic, Radioimmunoassay

Siemens Healthcare Diagnostics, Inc

ADVIA Centaur Folate (500 Tests) SMN 10325366, Cat No. 06891541 In vitro diagnostic use in the quantitative determination of folate in serum or red blood cells.

DateSep 10, 2013 (initiated Jul 25, 2013)
StatusTerminated
ReasonCalibration failures and/or significant negative shifts in quality control (QC) and patient results.
Root causeMaterial/Component Contamination
ActionSiemens sent an Urgent Medical Device Correction letter on July 26, 2013, to all affected customers via Federal Express. The letter identified the product, the problem, and the action to be taken by the customer. an Urgent Field Safety Notice was sent to all domestic customers via Federal Express. On July 25, 2013 this notice was sent to all OUS customers via hard copy, email, and/or fax as determined by each country's local regulations and procedures. These notices inform the customer that a portion of Folate ReadyPacks in kit lots ending in 218 are exhibiting calibration failures and/o
Quantity2840
DistributionWorldwide Distribution - USA including AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA,MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. Internationally to Canada and Mexico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K010050ADVIA CENTAUR AND ACS: 180 FOLATE IMMUNOASSAY
Bayer Diagnostics Corp., Feb 5, 2001