Recall Z-2099-2013 ยท HWC Screw, Fixation, Bone

Stryker Howmedica Osteonics Corp.

Stryker 1) Asnis III Cannulated Screw REF 604640 4.0x40mm TL 13.5mm 2) Asnis III Cannulated Screw REF 604642S 4.0x42mm TL 14mm, sterile Product Usage: The Asnis III Cannulated Screw System is intended for fracture fixation of small and long bones and of the pelvis. The system is not intended for spinal use.

DateAug 27, 2013 (initiated Jul 8, 2013)
StatusTerminated
ReasonStryker received a report that the length of the cannulated screw asnis III, article # 604640 Lot R11482, was incorrect.
Root causeProcess control
ActionStryker sent an Urgent Medical Device Removal letter dated July 8, 2013 via Fed Ex to all affected customers. The letter identified the affected product, product issues, potential hazards, risk mitigation and actions to be taken. Customers were instructed to return all affected products available to Stryer Osteosynthesis or contact Stryer customer service and complete and sign the enclosed Business Reply Form and fax a copy to: 1-865-251-3635. Questions can be directed to Stryker at 201-972-2100.
Quantity23 units
DistributionUSA Nationwide Distribution in the states of: FL. OH, TX, & VA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K000080ASNIS III CANNULATED SCREW SYSTEM
Howmedica Osteonics Corp., Apr 3, 2000