Recall Z-0070-2014 ยท LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

Zimmer, Inc.

Versys Beaded Fullcoat Femoral Stem (POR FULL-CT FEM ST 11X160MM; VERSYS BEADED FC 11X200MM 8ST; POR FULLCT FEM ST 11 LOW EXT; POR FULLCT FEM ST 11 EXT; POR FLCT FEM STEM LM 11X160MM; POR FULLCT LM FEM ST 11 EXT POR FULL-CT FEM ST 12X160MM VERSYS BEADED FC 12X200MM 8 ST POR FULLCT FEM ST 12 LOW EXT POR FLCT FEM ST 12.0X190 STR POR FULLCT FEM ST 12 EXT POR FULL-CT FEM ST LM 12X160MM POR FULLCT LM

DateOct 25, 2013 (initiated Aug 30, 2013)
StatusTerminated
ReasonThe low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.
Root causePackage design/selection
ActionAugust 30, 2013 Urgent Recall Notices were sent to Distributors, hospital risk managers and surgeons identifying the issue and directing consignees to make all affected personnel aware of the contents of the notification. The notice identified the products involved, the issue, possible adverse events associated with the issue and recommended that implants found with LDPE bag adhesion not use the product and return it to Zimmer sales representatives. Questions or concerns should be addressed to Zimmer customer call center at 1-877-946-2761 between the hours of 8 a.m. and 5 p.m. EST.
Quantity6380
DistributionWorldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Germany, Guatemala, Honduras, Hong Kong, India, Israel, Jamaica, Japan, Korea, Malaysia, Mexico, New Zealand, Nicarag
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K953337BETA HIP PROSTHESIS
Zimmer, Inc., Jan 22, 1996