ClearanceBrief
Recall Z-0198-2014 ยท HBC Motor, Drill, Electric

The Anspach Effort, Inc.

Pneumatic Systems Operating Manual (part numbers 19-0005) utilized with the Pneumatic Motor Systems including: BlackMax Motor Systems, MicroMax Systems and XMax Motor Systems all models and serial numbers. Product Usage: Indication for use is cutting and shaping bone.

DateNov 13, 2013 (initiated May 6, 2011)
StatusTerminated
ReasonAnspach Effort, Inc. Palm Beach Gardens, FL initiated a correction of its medical Device labeling for its Pneumatic Systems Operating Manual, Cutting Burr Chart and Product Catalog due to a recent review of their regulations and compliance requirements that determined the claim that the device is capable of dissection of biometals, bioplastics and cement has been verified, but is is not part of th
Root causeLabeling Change Control
ActionThe firm, Anspach Effort, Inc., sent an "Urgent Medical Device Labeling Correction" letter dated June 27, 2011 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to do the following: 1) Destroy or delete any items in their inventory immediately and replace with the attached revised Operating Manual, Product Catalog and/or Cutting Bur Chart. 2) Complete and return the attached Customer Reply Form indicating disposition of the labeling materials and confirming their receipt of this letter, and fax to 1-800-327-6661. 3) If they dist
Quantity2,020
DistributionWorldwide Distribution: USA Nationwide and the countries of: Egypt, India, Germany, Turkey, Canada, Saudi Arabia, Greece, United Kingdom, Cayman Islands, Kazakhstan, Republic of Korea, Taiwan, Hong Kong, China, Lebanon, Argentina, Denmark, Belgium, Netherlands, Mexico, France, Australia, Panama, Ita
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011444ANSPACH EMAX DRILL SYSTEM
The Anspach Effort, Inc., Aug 8, 2001