ClearanceBrief
Recall Z-1981-2013 ยท JDG Prosthesis, Hip, Femoral Component, Cemented, Metal

Zimmer, Inc.

Moore Hip Prosthesis Long Fenestrated Stem, several head Diameter, 203 mm Stem length. Used to seat the prosthesis with or without bone cement.

DateAug 17, 2013 (initiated Jul 16, 2013)
StatusTerminated
ReasonZimmer recently conducted a review of historic packaging validations completed. Based on this review, it was determined that a subset of product packaged at a single Zimmer facility required testing to ensure that previously packaged product from this facility met the established requirements for sterile barrier integrity. This testing determined that a single packaging tray configuration, which
Root causePackaging process control
ActionZimmer is removing the devices from the field including distributor and hospital inventories. A notification was issued to known users of the product on 07/16/2013.
Quantity1,710 units are in distribution
DistributionWorldwide Distribution-USA (nationwide) including the states of CT, FL, GA, IL, IN, KS, KY, MA, MI, MN, MT, NC, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VI, and WI, and the countries of Singapore, Canada, Mexico, Arab Emirates, Saudi Arabia, and Hong Kong.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.