Recall Z-0090-2014 ยท IZL System, X-Ray, Mobile

Carestream Health, Inc.

DRX Revolution Mobile X-Ray System, Model DRXR-1, Verona Street Rochester, NY 14608 Catalog Numbers 1019397, 1023415, 1023423 --- Made in USA by Carestream Health, Inc. 150 The device is designed to perform radiographic x-ray examinations on all pediatric and adult patients, in all patient treatment areas.

DateOct 31, 2013 (initiated Jul 26, 2013)
StatusTerminated
ReasonDuring an audit, Carestream discovered a DRX-1 System that captured an image which displayed an artifact. This issue was discovered in manufacturing. It occurs when the DRX product console software is updated to v5.6b software and the detector is not calibrated before use.
Root causeSoftware design
ActionCarestream sent a Urgent-Medical Device Recall letter dated July 26, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Carestream Health Inc. recently discovered an image artifact issue due to a detector not being recalibrated with the 5.6B software upgrade. As our installation instructions did not require the user to recalibrate during an upgrade, there is a possibility your detector could do the same. When this occurs an image artifact could be evident on images captured in the Extended Exposure mode. These arti
Quantity23 units installed total (11 domestically & 12 internationally)
DistributionWorldwide Distribution - US (nationwide) including the states of New York, Massachusetts, Pennsylvania, Illinois, Texas, California and Louisiana., and the countries of Australia, Canada, Finland, Italy, Switzerland and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120062DRX-REVOLUTION MOBILE X-RAY SYSTEM
Carestream Health, Inc., Apr 13, 2012