Recall Z-1979-2013 ยท LWS Implantable Cardioverter Defibrillator (Non-Crt)

Boston Scientific CRM Corp

Boston Scientific, ENERGEN DR ICD, Model E143. The device is an Implantable Cardioverter Defibrillator.

DateAug 16, 2013 (initiated Jul 15, 2013)
StatusTerminated
ReasonBoston Scientific CRM manufacturing quality system recently discovered test artifacts (related to shock charge time) in a small number of implantable defibrillators (ENERGEN DR ICD model E143) which require further investigation. While these devices successfully passed all manufacturing tests and met all requirements, they showed a slightly longer charge time than other devices during a manufactu
Root causeUnder Investigation by firm
ActionConsignee was sent on 7/15/13 a Boston Scientific "Medical Device Retrieval" letter dated July 12. 2013. The letter described the product problem and informed the consignee that they will replace the product removed from their inventory. For further questions, contact Boston Scientific Technical Services at 1 800 227-3422.
Quantity1 (4 devices were implanted and are not part of this action)
DistributionUS distribution in the state of Tennessee.
FDA recordOpen on fda.gov