Recall Z-2110-2013 ยท MKQ Processor, Cervical Cytology Slide, Automated

Becton Dickinson & Co.

PrepStain AG. For use in the screening and detection of cervical cancer.

DateAug 28, 2013 (initiated Jul 8, 2013)
StatusTerminated
ReasonPrepStain system rack setting was incorrectly entered on one PrepStain system during a routine preventive maintenance visit by a BD Field Service Representative.
Root causeSoftware design
ActionEx-US BD sites were notified via email and teleconference, and will communicate directly with customer sites in their region as required by local regulation. All US customers were initially contacted via telephone and letters were mailed.
Quantity890 instruments
DistributionWorldwide Distribution-USA (nationwide) and Puerto Rico and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.