Recall Z-1911-2013 ยท LTW Susceptibility Test Cards, Antimicrobial
Biomerieux Inc
bioMerieux Vitek 2 Systems Software version 6.01 software, used with the VITEK 2 automated system. Kit, UPDT V2S 6.01 EN. Part number: 414560. The product is used with the VITEK 2 automated system which consists of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast.
| Date | Aug 6, 2013 (initiated Jun 14, 2013) |
| Status | Terminated |
| Reason | A product issue has been identified for customers using software version 6.01 and are interfaced with a Laboratory Information System (LIS). During the update of software from version 5.xx to 6.01, the Bid-directional Computer interface (BCI) antibiotic codes are not updated properly for Cefovecin, Prulifloxacin and Ceftaroline. For the referenced antibiotics, the firm code in the reference table |
| Root cause | Software change control |
| Action | The firm, bioMerieux, sent an "Urgent Product Correction Notice" letter dated June 14, 2013 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to ensure the Urgent Product Correction Notice and the Instructions for modifying BCI Translation Codes are distributed to all appropriate personnel in your organization; follow the instructions in the attached file (Instructions for modifying BCI Translation Codes) to resolve the discrepancy associated with Cefovecin, Prulifloxacin and Ceftaroline in the translation table and complete and |
| Quantity | 12,384 total kits |
| Distribution | Worldwide distribution: US (nationwide) and countries of: Argentina, Australia, Austria, Bahrain, Belarus, Benelux, Brazil, Bulgaria, Canada, Chile China, Costa Rica, Croatia, Czech Republic, Dominican Republic, Dutch Antilles, Ecuador, Egypt, France, Germany, Hong Kong, Hungary, India, Indonesia, I |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.