Recall Z-1956-2013 ยท HAW Neurological Stereotaxic Instrument

Medtronic Navigation, Inc.

StealthStation S7 Framelink Software version 5.4. Aid for locating anatomical structures and planning surgical trajectories in open and percutaneous procedures.

DateAug 13, 2013 (initiated Jul 15, 2013)
StatusTerminated
ReasonMedtronic Navigation, Inc. has initiated a field correction to their Framelink software version 5.4 which can be used on Medtronic Navigation StealthStation 57 and i7 systems or a Medtronic Planning Station due to the potential of discrepancy between the exam image and the displayed measurement or overlay under certain conditions of use.
Root causeSoftware design
ActionA Field Safety Notice was sent on 7/15/13 to users informing them that the Framelink software version 5.4 has been revised to correct the software anomalies described in this letter. A new version will be installed on their system(s) by a Medtronic Navigation representative. Until the software upgrade has been installed on their system, do not use high resolution examinations (greater than 256 x 256). If users are not sure of the exam resolution being used, firm recommend contacting the hospital imaging department or local Medtronic Navigation representative for assistance. Medtronic represe
Quantity572
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K992927STEALTHSTATION TREATMENT GUIDANCE PLATFORM
Surgical Navigation Technologies, Inc., Dec 29, 1999