Recall Z-2210-2013 ยท JAK System, X-Ray, Tomography, Computed

Philips Medical Systems (Cleveland) Inc

Brilliance CT 6, M/N 728256; Brilliance CT 16 M/N 728246; Brilliance CT 40, M/N 728235 & Brilliance CT 64 M/N 728231, Philips Healthcare, Cleveland, OH 44143. Intended to produce cross-sectional images of the body by computer reconstruction.

DateSep 16, 2013 (initiated Jul 5, 2013)
StatusTerminated
ReasonThe patient support may move in an unintended manner if the footswitch cover impinges on the footswitch.
Root causeComponent design/selection
ActionPhilips Healthcare sent an "URGENT-MEDICAL DEVICE CORRECTION" letter dated June 4, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact the Customer Care Center at 1-800-722-9377, option 5 for information concerning this issue.
QuantityM/N: 728256, 4 units; M/N: 728246, 89 units; M/N: 728235, 9 units & M/N: 728231, 207 units
DistributionWorldwide Distribution-USA (nationwide) including DC and the states of AL, AR, AZ, CA, CO, DE, FL, GA, IN, KY, MA, MD, MI, MN, MS, NC, NM, NV, NY, OH, OK, PA, SC, TN, TX, VT, WI & WV, and the countries of Australia, Austria, Bosnia & Herzegovina, Chile, China, Croatia, Denmark, Equatorial Guinea, Fr
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012009MX8000 IDT CT SYSTEM
Marconi Medical Systems, Inc., Sep 13, 2001
K033326PHILIPS PLUS COMPUTED TOMOGRAPHY SYSTEM
Philips Medical Systems, Inc., Oct 29, 2003