Recall Z-1729-2013 ยท FMK Single Use Only Blood Lancet With An Integral Sharps Injury Prevention Feature

KooJoo Trading Company

Safety Lancets - Odyssey, Button Activated; Tri- Bevel Tip for Virtually Painless Sampling; Needle automatically retracts, No lancing device required. 26G - Single Use Only, Made In Korea Distributed by Hummingbird Medical, Savannah, GA 100 count Boxes; 40 boxes per case Blood collection by penetrating skin with needle.

DateJul 13, 2013 (initiated Jul 3, 2013)
StatusTerminated
ReasonLancets were not sterilized within the dose range established by sterilization validation.
Root causeProcess control
ActionThe firm, KooJoo, sent a "MEDICAL DEVICE VOLUNTARY T RECALL" letter dated July 3, 2013 via first class mail to its customer. The letter described the product, problem and actions to be taken. The customer was instructed to identify and hold all safety lancets you have in inventory that was received before June 2013; notify your retail customers of this product recall and complete and return the attached Customer Reply Form via email to heegonkim@hotmail.com and also return form along with any unused/undistributed product for replacement. If you have further questions, please email heegonkim@h
Quantity7, 712, 000 pieces (1928 cartons)
DistributionUS Distribution GA only.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.