Recall Z-1782-2013 ยท IZI System, X-Ray, Angiographic
Siemens Medical Solutions USA, Inc
ARTIS zee and zeego x-ray, angiographic system with Software Artis VC20x/VC21A that have Patch 7 and a 3D and DSA license installed. Angiographic x-ray system
| Date | Jul 22, 2013 (initiated Jun 26, 2013) |
| Status | Terminated |
| Reason | Siemens issued a customer safety advisory notice and field safety corrective action for a potential issue on Artis systems with Artis VC20x/VC21A that have Patch 7 and a 3D and DSA license installed. The issue affects-DSA rotational acquisition. The restriction is related to any rotational 3D-acquisition program with real-time subtraction (e.g. 3D DSA). During the fill phase of the rotational aqui |
| Root cause | Software Design Change |
| Action | Siemens sent a Customer Safety Advisory Notice dated June 26, 2013, to all affected customers. The letter identified the product, the problem, and the action needed to be taken by the customer. Customers were strongly recommend to carefully evaluate one of the following options before running 3D rotational acquisition programs that include real-time subtraction until a patch is installed on their system: 1) position the monitor in a way such that contrast advancement can be monitored during the fill phase on the live monitor in the control room. 2) For applications not requiring actions |
| Quantity | 29 |
| Distribution | Nationwide Distribution and Puerto Rico |