ClearanceBrief
Recall Z-1780-2013 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Philips Healthcare Inc.

Radiohead Access Point 1.4 GHz Radiohead Access Point Firmware versions: C.00.04 & C.00.05 Part Number: 989803171211 Product Usage: IntelliVue Smart-hopping 1.4 GHz Access Point provides wireless connectivity between IntelliVue devices (listed below) and the Philips IntelliVue Information Center. MX40 Indicated for use by health care professionals whenever there is a need for monitoring the ph

DateJul 20, 2013 (initiated Jul 8, 2013)
StatusTerminated
ReasonIntelliVue Smart-hopping 1.4 GHz Access Point may experience unexpected signal loss of several minutes impacting wireless connectivity
Root causeSoftware Design Change
ActionPhilips sent an Urgent Medical Device Correction letter dated July 2013 to affected customers. The letter informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer. Philips will perform a firmware upgrade (version C.00..07 for all affected customers at no cost. Philips is asking customers to follow the Action to be Taken by Customer/User section of the Urgent Medical Device Correction notification/Field Safety Notice. If you need any further information or support concerning this issue, please contact Philips Customer S
Quantity20105 units
DistributionUSA Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041741M3290A INTELLIVUE INFORMATION CENTER SOFTWARE RELEASE F.0 AND M4840A PHILIPS TELEMETRY SYSTEM II INCLUDING M4841A PATIEN
Philips Medical Systems, Inc., Jul 21, 2004