Recall Z-1750-2013 ยท IZI System, X-Ray, Angiographic

Siemens Medical Solutions USA, Inc

Siemens AXIOM Artis zee/zeego systems equipped with generator A100 Plus Product Usage: Angiographic x-ray system

DateJul 17, 2013 (initiated Jan 31, 2013)
StatusTerminated
ReasonDuring the course of product monitoring, Siemens became aware of a tolerance problem in the power supply of the generator control above a specific value, thus, the generator may fail.
Root causeDevice Design
ActionSiemens sent an Urgent Customer Information letter to their consignees. The letter identified the affected product, problem , update Instructions and actions to be taken. Siemens began to equip affected systems with a modified board assembly in January, 2013.
Quantity23
DistributionUnited States Nationwide Distribution - including the states of: CA, CO, FL, IA, IL, IN, MA, MD, MI, NC, NY, PA, TX, VA and WA and Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090745ARTIS ZEE / ZEEGOS WITH SW VC14
Siemens Medical Solutions USA, Inc., Jun 18, 2009