Recall Z-1875-2013 ยท IZL System, X-Ray, Mobile

Siemens Medical Solutions USA, Inc

Mobiletta Mira Mobile X-ray System

DateAug 5, 2013 (initiated May 2, 2013)
StatusTerminated
ReasonSiemens issued a Field Safety Alert about the potential risk to users when operating the Mobilett Mira. When a Mobilett Mira is regularly exposed to extensive mechanical strain (driven over high thresholds for doors or elevators), it is possible that one of its main wheel shafts could break because of material fatigue, resulting in the chassis falling on the ground on the affected wheel side. This
Root causeDevice Design
ActionSiemens sent a Field Safety Notice dated May 2. 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. Customers were instructed to : 1) Never step or sit on the unit or otherwise put substantial weight on it. 2) Avoid moving the unit over high thresholds for doors or elevators where possible, or do so carefully and slowly. 3) Make sure that no person had his/her extremities under the chassis, also when the unit is not moving. Update instructions were also released to alert customers of the potential risk and to
Quantity39
DistributionUS Distribution including the states of IA, SD, NC, WI, OH, NE, TX, KS, ID, MI, GA, AR and IL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111912MOBILETT MIRA
Siemens Medical Solutions USA, Inc., Sep 19, 2011