Recall Z-1795-2013 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Siemens Medical Solutions USA, Inc

Axiom Aristos MX Solid State Flat Panel Digital Imager, Model number 5895003. The device is used for diagnostic X-ray imaging.

DateJul 24, 2013 (initiated Jun 5, 2013)
StatusTerminated
ReasonSiemens issued a Field Safety Notice about the potential hazard to patients or operators when using the function "table top float" of the AXIOM Aristos MX. When repositioning a patient lying on the table top and releasing the table top, it is possible that the fingers of the patient or operator are between the table top and the detector tray and if the table top is moved, the fingers can be squee
Root causeDevice Design
ActionOn 6/5/13, consignees were notified by letter of a "Field Safety Notice" and received instructions about how to avoid the potential risk of this issue while using the device. Siemens also rolled out a modification of AXIOM Aristos MX that will avoid these potential injuries in the future.
Quantity401
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K983732THORAX FD AND MULTIX FD
Siemens Medical Solutions USA, Inc., Apr 21, 1999