ClearanceBrief
Recall Z-1789-2013 ยท KPS System, Tomography, Computed, Emission

GE Healthcare, LLC

Elscint APEX Helix, dual-head, multi-purpose, Slip-Ring, digital gamma camera Product Usage: The intended use of the Quasar system is to perform general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition typed and optional imaging features designed to enhance image quality in Onco

DateJul 24, 2013 (initiated Jun 13, 2013)
StatusTerminated
ReasonGE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System.
Root causeDevice Design
ActionGE sent an Urgent Medical Device Correction letter dated June 17, 2013 to all affected customers. The letter identified the affected product, recommended that qualified service personnel maintain the equipment and that Preventative Maintenance procedures are executed according to labeling. Also the Safety Chapter Sections should be re-reviewed with personnel to ensure proper operation of the equipment. GE notified customers again on July 03, 2013, via Urgent Medical Device Recall letter (including confirmation of delivery for US customers) and follow-up telephone call. Customers were instr
Quantity50
DistributionWorldwide Distribution - US Nationwide including Guam, Puerto Rico and the countries of : Afghanistan, Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, G
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K914807APEX HELIX - A
Elscint, Inc., Feb 27, 1992