Recall Z-1830-2013 ยท KHO Fluorometer, For Clinical Use
Tosoh Smd Inc
AIA Analyzer Pipette Tips used on Tosoh AIA-2000, AIA-1800, AIA-900 and AIA-600II Analyzers, 1000 tips/bag AIA-Analyzer Tips are disposable pipette tips packaged in bulk quantities and used on Tosoh Bioscience AIA Analyzers for lab analysis of various immunoassays.
| Date | Jul 30, 2013 (initiated May 15, 2013) |
| Status | Terminated |
| Reason | Visual inspection of the pipette tips indicated that some of the tips were not uniform in size/length which was causing error messages on the analyzers. |
| Root cause | Nonconforming Material/Component |
| Action | Tosoh Bioscience sent an Urgent Recall Notification letter dated May 15, 2013. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to respond on the amount of product that was destroyed by completing the attached form and faxing it to 614-317-1941. Customers with questions were instructed to call Tosoh Technical Support at 1-800-248-6764. For questions regarding this recall call 614-317-1909. |
| Quantity | 11,309 bags |
| Distribution | Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WY. Internationally to Canada, Japan, Mexico, Chile, and Uruguay. |
| FDA record | Open on fda.gov |