Recall Z-2090-2013 ยท KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component

Zimmer, Inc.

Free Lock Femoral Hip Fixation System Compression Lag Screw . Several lengths (55 to 150 mm) and diameters (12.7 and 15. 8 mm). Item codes starting with 00-1181-05, 00-1181-10 and 00-1181-15. ZIMMER Compression Hip Screw, orthopedic implant.

DateAug 26, 2013 (initiated Jun 19, 2013)
StatusTerminated
ReasonSingle packaging configuration used for the VERSA-FX-II Femoral Fixation System failed the leak test compromising the sterility of the product.
Root causePackaging process control
ActionThe firm, Zimmer, sent an "URGENT MEDICAL DEVICE NOTIFICATION" letter dated July 11, 2013 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to inspect the device according to the insert instructions; not use the device if damage to the corner of the inner tray is observed, and contact the firm for return and questions. Customers should also keep this notification in mind and inform Zimmer of product with this condition using zimmer.per@zimmer.com. If after reviewing this notification you have questions or concerns, please contac
Quantity85, 794 units in total
DistributionWorldwide distribution: US Nationwide (including PR and USVI), and countries of: Israel, Argentina, Brazil, France, Mexico, Saudi Arabia, Beirut, China, Korea, San Salvador, Honduras, Canada, Australia, Switzerland, Germany, Arab Emirates, UK, Belgium, Egypt, Spain, France, Italy, Iran, Lebanon, Jor