Recall Z-1837-2013 ยท BXB Exerciser, Powered
Accelerated Care Plus Corporation
Omnicycle Elite; US Importer and Distributor: Accelerated Care Plus, Reno NV. Manufactured by Medica Medizintechnik GmbH, Hochdorf, Germany. The product is indicated for powered exercise equipment intended for medical purposes.
| Date | Jul 31, 2013 (initiated Jun 12, 2013) |
| Status | Terminated |
| Reason | An anomaly in the firmware that may cause Omnicycle Elite to operate at a speed other than what was previously set when in reverse mode. |
| Root cause | Software design |
| Action | Letters were issued on June 12, 2013 that included a memory stick with self loading software so consumers can update the device. Letters were sent by trackable method. Customers with questions or concerns should call Customer Support at 800-350-1100. |
| Quantity | 1001 units |
| Distribution | Nationwide Distribution. |
| FDA record | Open on fda.gov |