Recall Z-1661-2013 ยท JOW Sleeve, Limb, Compressible

DJO, LLC

VenaFlow Elite 30B-B and 30BI-B Product Usage: The VenaFlow Elite System is a compression pump used for the prevention of Deep Vein Thrombosis.

DateJul 3, 2013 (initiated May 28, 2013)
StatusTerminated
ReasonDJO, LLC has recently identified a product safety issue with the battery assembly for the VenaFlow 30B-B and 30 BI-B products. There is a potential for battery leakage which could result in thermal meltdown (fire).
Root causeDevice Design
ActionDJO sent a Customer Notification letter dated May 28, 2013 to Domestic customers via USPS certified return receipt and to International customers via UPS on June 14, 2013. The Customer Notification informed customers of the product safety issue, product description, and instructions to follow. DJO plans to install replacement batteries at no charge. For questions contact your local DJO Customer Service Representative.
Quantity2491
DistributionWorldwide Distribution - USA Nationwide and countries of: Australia, Belgium, China, Guatemala CentroAmeria, France, Hong Kong, India,Italy, Korea, Lebanon, Malaysia, Morocco, Pakistan, Paraguay, The Netherlands, Romania, Singapore, Spain, Sweden, Thailand, Turkey, United Arab Emirates, United Kingd
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K091700AIRCAST VENAFLOW ELITE SYSTEM
Djo, LLC, Jul 9, 2009