Recall Z-1694-2013 ยท MUE Full Field Digital, System, X-Ray, Mammographic

Siemens Medical Solutions USA, Inc

Siemens Mammomat Inspiration - full-field digital mammography Product Usage: Mammography

DateJul 11, 2013 (initiated Feb 28, 2013)
StatusTerminated
ReasonFirm became aware of an unintended behavior when using the Mammomat Inspiration. The possibility exists that a malfunction may occur during Stereo or Tomo acquisitions if the tomo clutch was incorrectly adjusted. Although the possibility is rare, the system could move while a patient's breast is compressed resulting in a possible bruise. Firm released Update Instructions XP031/12/S for this iss
Root causeDevice Design
ActionSiemens issued an Update Instructions XP031/12/S by letter to the affected customers. For affected sites, a Siemens Service Engineer will inspect the unit and make any adjustments needed. Following the inspection and adjustments, a proper calibration will be completed. For questions call 610-448-3237 and 610-446-4634.
Quantity3
DistributionUnited States Nationwide Distribution in the states of ND, OH and SC.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122286MAMMOMAL INSPIRATION
Siemens Medical Solutions USA, Inc., Feb 22, 2013