Recall Z-1887-2013 ยท DTK Filter, Intravascular, Cardiovascular

Cordis Corporation

Cordis OPTEASE Retrieval Inferior Vena Cava (IVC) Filter***REF Cat. No.***466F220A and 466F220B***Vena Cava Filter and Introduction Kit***(BRITE TIP Catheter Sheath Introducer and Angiographic Vessel Dilator)***Cordis***a Johnson and Johnson Company***ENDOVASCULAR*** Label No. SB 10247519.2/***LB466F220AO.5***omoraoro/2012.050794***Do not resterilize***. Product Usage: The OPTEASE Vena Cava Filt

DateAug 7, 2013 (initiated Mar 29, 2013)
StatusTerminated
ReasonCorrection to provide clarification and additional labeling to minimize likelihood on implanting filter backwards.
Root causeLabeling design
ActionCordis Corporation sent an "URGENT Medical Device Correction" notification to their consignees in the US by letter dated April 3, 2013 and an e-mail was sent to the Canadian consignees. The letter identified the affected product, problem and actions to be taken. Consignees were instructed to read the description and recommendations sections carefully, then sign and return the enclosed Acknowledgement Form directly to Cordis to their fax number(908) 429-8287 or Scan and e-mail to MiamiQA@crdus.jnj.com. The letter states to pass on the notice to anyone in the consignees' facilities that need t
Quantity33,000 units
DistributionWorldwide Distribution - USA Nationwide in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, Ny, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, including Hawaii and Puerto Rico and the countries
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K034050CORDIS OPTEASE VENA CAVA FILTER, MODELS 466-F220A AND 466-F220B; RETRIEVABLE CATHETER, MODEL 466-C220F
Cordis Corporation, Mar 22, 2004
K091077OPTEASE VENA CAVA FILTER AND RETRIEVAL CATHETER
Cordis Corp., A Johnson & Johnson Co., Feb 4, 2010