Recall Z-1697-2013 ยท HRY Prosthesis, Knee, Femorotibial, Semi-Constrained, Cemented, Metal/Polymer

Biomet, Inc.

REF 154202 Oxford Fixed Bearing Partial Knee Cemented Lateral Tibia Size A5 ARCOM UHMWPE / CO-CR-MO Alloy Only for use in Lateral compartment Sterile R The Oxford Fixed Bearing Knee is a Partial knee replacement system consisting of a femoral component and a fixed bearing tibial component for use in the lateral compartment of the knee. The Oxford Fixed Lateral bearings are indicated for use in th

DateJul 11, 2013 (initiated Jun 7, 2013)
StatusTerminated
ReasonBiomet investigation found the design of the tibial bearing creates the potential for anterior impingement between the bone and the poly bearing component in terminal extension.
Root causeDevice Design
ActionThe firm, Biomet, sent an "URGENT MEDICAL DEVICE RECALL NOTICE" letter dated June 7, 2013 to its consignees/customers. The letter identified the device being recalled, the reason for the recall, potential adverse events associated with the recall and actions to be taken. The consignees/customers were instructed immediately locate and remove the identified device(s) listed in the letter; carefully follow the instructions on the enclosed "FAX Back Response Form"; complete and return the FAX Back Response Form to 574-372-1683 prior to return of product within three (3) business days; return Goo
Quantity28
DistributionWorldwide distribution: USA including NC, IN, PA, and OH; and countries of: Canada and Netherlands.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122277OXFORD FIXED LATERAL BEARING PARTIAL KNEE REPLACEMENT (OXFORD TIBIAL TRAYS A-F) IXFORD PARTIAL KNEE REPLACEMENT SYSTEM
Biomet Manufacturing Corp, Oct 22, 2012