Recall Z-1572-2013 ยท FGY Cannula, Injection

Maquet Cardiovascular Us Sales, Llc

MAQUET GETINGE GROUP Arterial Cannula OD Used to introduce blood and sterile solutions into the systemic circulation during pediatric extracorporeal circulation procedures.

DateJun 21, 2013 (initiated May 14, 2013)
StatusTerminated
ReasonThe specified part numbers and lot numbers of Maquet Pediatric Arterial Cannula Packs may contain an incorrectly sized connector.
Root causePackaging process control
ActionMaquet sent an "Urgent Product Recall"/"Recall - Device Corrective Action Response" forms dated May 14, 2013, via Federal Express to all affected customers. The notification identifies affected product; informs customers of the nature of the issue; advises of potential adverse patient consequences that may occur; and advises customers to follow instructions on the "Recall-Device Corrective Action Response" forms included as part of the notification. If you have any further questions please contact your local MAQUET representative or call Customer Service at (888) 627-8383.
Quantity100 units (60 Domestic; 40 International)
DistributionWorldwide Distribution - US Distribution including the states of Delaware, Illinois, Minnesota , New York., Ohio and Pennsylvania and the countries of Germany, Turkey and France.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.