Recall Z-1586-2013 ยท MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)

Medrad Inc

DC power cable (p/n 3010556) supplied with the MEDRAD(R) Veris MR Vital Signs Monitor: Product Usage: Monitor, Physiological, Patient

DateJun 25, 2013 (initiated Apr 15, 2013)
StatusTerminated
ReasonThe firm is recalling these power cables due to a latent design reliability issue and the potential for shorting which can result in heating/melting of the cable jacket.
Root causeDevice Design
ActionMedrad Inc sent an Urgent Medical Device Component Recall letter dated April 15, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter states that replacement products will be available in June and shipped in the coming months along with additional instructions for returning the affected product. Customers were instructed to complete the enclosed recall response form and fax it back to Bayer at 412-406-0942. For questions contact customer support team at 1-877-229-3767 and select option 2.
Quantity1,066
DistributionWorldwide Distribution - US Nationwide including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY and the countries of: AE, AU, AZ, BN, B
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K0425698600 VITAL SIGNS MONITOR
Criticare Systems, Inc., Sep 29, 2004